产品信息
货号:1006
品名:HTD 96b – Complete Unit
品牌:HTDialysis
原文
Q:The customer who purchased this product wants to know if there is a matching cleaning detergent available for purchase. If not, they would like to know your recommended cleaning method or whether any detergent can be used.
A:After each use the HTD dialysis block should be cleaned thoroughly with a non-ionic detergent. Disassemble the dialysis block and soak overnight in a 2-liter beaker containing a 2% solution of Micro 90 (VWR international, Catalog # 21830-416). Rinse the Teflon bars in the 2-liter beaker followed by holding each Teflon bar under a running distilled water stream. Care must be taken to ensure that all detergent has been rinsed from the unit as residual detergent could compromise future binding studies. The Teflon block may also be disinfected with a 10% v/v solution made using commercial Clorox bleach which is 5.25% sodium hypochlorite. The final solution would this be 0.525% sodium hypochlorite solution. Never use any abrasive or brush for cleaning the Teflon blocks as they will cause micro-striations and prevent effective sealing.We do not sell a cleaning detergent.
译文
问:购买了该产品的客户想了解是否有配套的清洁剂可供购买。如果没有,他们希望获知您推荐的清洁方法,或是询问是否可以使用任何清洁剂。
答:每次使用后,HTD透析块必须用非离子型洗涤剂清洗足够。请按以下步骤操作:
1. 将透析块拆开,浸泡在装有2% Micro 90溶液(VWR国际,目录号#21830-416)的2升烧杯中过夜。
2. 在2升烧杯中漂洗特氟龙条,随后将每根特氟龙条置于流动蒸馏水下冲洗。
⚠️ 务必确保冲净所有洗涤剂残留,否则可能影响后续结合实验的准确性。
特氟龙块亦可使用含5.25%次氯酸钠的商用Clorox漂白剂配制10%(体积比)溶液进行消毒,最终浓度为0.525%次氯酸钠。
❗ 严禁使用任何研磨剂或刷子清洁特氟龙块,以免产生微细划痕导致密封失效。
❗我们不销售清洁剂。
更多产品信息,请联系中国区域代理商:上海起发实验试剂有限公司
(注:本文内容基于品牌公开资料及行业常规信息整理,具体以品牌信息文档为准。)
电话
微信